E-20A-H-O / E-20AJ-H-O User Manual V2.1
13 / 36
Eng
lis
h
- Use the accompanying tube and mask provided by BMC.
- Do not wear the mask for more than a few minutes while the device is not operating.
- Use only masks with vent holes. Do not block or try to seal the vent holes in the exhalation
port.
11.5 Using Oxygen with the Device
Oxygen may be added at the mask connection. Please observe the instructions listed below
when using oxygen with the device.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.
• Turn on the device before turning on the oxygen. Turn off the oxygen before turning off the
device. Explanation of Warning: When the device is turned off, but the oxygen flow still exists,
oxygen may accumulate within the device's enclosure and pose a fire hazard. Turning off the
oxygen before turning off the device will prevent oxygen accumulation in the device and reduce
the risk of fire. This warning applies to most CPAP devices.
• Oxygen supports combustion. Keep the device and the oxygen container away from heat,
open flames, any oily substances, or other sources of ignition. DO NOT smoke in the area near
E-20A or the oxygen container.
• Sources of oxygen should be located more than 1 m from the device.
• When using oxygen with this system, a Pressure Valve must be placed in-line with the patient
circuit between the device and the oxygen source. The pressure valve helps prevent the
backflow of oxygen from the patient circuit into the device when the unit is off. Failure to use
the pressure valve could result in a fire hazard.
• Do not connect the device to an unregulated or high pressure oxygen source. The pressure of
oxygen source does not exceed the work pressure of the device.
11.6 Inserting the SD Card
(
Only for the device that
equipped with SD card)
Insert the SD card into the SD Card Slot, as shown in Fig. 11-6.
SD Card Slot
Fig. 11-6
If the SD card is inserted correctly, a symbol indicating correct insertion will appear in the Main
Interface on the screen of the device, as shown in Fig. 11-7.
Summary of Contents for E-20A-H-O
Page 1: ......
Page 2: ...English User Manual RESmart GII Auto CPAP System E 20A H O E 20AJ H O 0123...
Page 41: ...Espa ol Manual del Usuario RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 80: ...Italiano Manuale Utente RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 118: ...390400...