E-20A-H-O / E-20AJ-H-O User Manual V2.1
16 / 36
Eng
lis
h
Communications Port.
The model of the SpO
2
Kit is KS-CM01. The SpO
2
Kit is calibrated to display FUNCTIONAL
OXYGEN SATURATION.
Attach its sensor to the patient's index finger or any other finger.
The sampling rate of the SpO
2
signal is about 50 Hz, and the update rate of the frame is 1 Hz.
The value of SpO
2
and PR is calculated by the average of the former eight pulse waveforms.
If the SpO
2
Kit is in an abnormal state, the value of SpO
2
will be blank.
The screen of the main device then displays the Main Interface shown in Fig. 11-14 (only
applies to E-20A-H-O), or the Main Interface shown in Fig. 11-15 (only applies to E-20AJ-H-O).
The patient's blood oxygen saturation and pulse rate can be clearly seen during the course of
therapy.
Fig. 11-14
Fig. 11-15
WARNINGS!
• Change the measurement point regularly according to the patient's conditions after prolonged
use. Change the measurement point, check the patient's skin integrity and circulatory
conditions, and make the right adjustments at least every three hours.
• Excessive ambient light, excessive motion, use of intravascular dyes, poorly perfused finger,
extreme finger sizes or improper placement may degrade the SpO
2
Kit's performance or affect
the accuracy of the measurement.
• Nail polish or false nails should be removed before the finger sensor is used, or it may cause
erroneous measurements results.
• Overly low blood pressure, overly low systolic blood pressure, severe anemia, or hypothermia
may cause erroneous measurements results.
• The SpO
2
Kit is designed for use with this device only.
• Verify the compatibility of the device and SpO
2
Kit before use, otherwise it may cause injury to
the patient.
• Misapplication of a SpO
2
Kit with excessive pressure for prolonged periods can induce
pressure injury.
Summary of Contents for E-20A-H-O
Page 1: ......
Page 2: ...English User Manual RESmart GII Auto CPAP System E 20A H O E 20AJ H O 0123...
Page 41: ...Espa ol Manual del Usuario RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 80: ...Italiano Manuale Utente RESmart GII Auto CPAP Sistema E 20A H O E 20AJ H O 0123...
Page 118: ...390400...