background image

InH2

®

 

Heated Humidifier

 User Manual V2.7

 

1/13 

English 

Warnings & Cautions 

IMPORTANT! 

• Read all instructions before using the humidifier. 
• Use only with BMC RESmart

®

 devices whose instructions specify the use of 

this humidifier. 

CAUTIONS! 

• Indicates the possibility of damage to the device. 
• US federal law restricts this device to sale by or on the order of a physician. 
• If fluids are spilled onto the humidifier platform, unplug the power cord from 
the AC wall outlet and allow the humidifier platform to drain and dry before 
using. 
• Take precautions to protect furniture from water damage. 

WARNINGS! 

• Indicates the possibility for injury to the user or the operator. 
• Use the humidifier only for its intended use as described in this manual. 
• Use only accessories recommended by BMC. 
• Never operate the humidifier if any of the parts are damaged, if it is not 
working properly, or if the humidifier has been dropped or mishandled. Do not 
use the humidifier if the water chamber is leaking or damaged in any way. 
Have any damaged parts replaced before continuing use. 
• Never touch the heater plate unless the humidifier is unplugged and the plate 
has cooled down. 
• This equipment is not suitable for use in the presence of a flammable 
anesthetic mixture with air or with oxygen or nitrous oxide. 
• Periodically inspect the power cord for signs of wear or damage. Replace if 
necessary. 
•Do not operate the device in direct sunlight or near a heating appliance 
because these conditions can increase the temperature of the air coming out 
of the device.   
• When humidifier is used outside the specified ambient temperature range or 
humidity range, the performance of humidifier will be compromised. 

Summary of Contents for INH2

Page 1: ...English InH2 Heated Humidifier User Manual 0123...

Page 2: ...er is only used for single patient and must not be re used on another person This is to avoid the risk of cross infection The InH2 Heated Humidifier is not intended for use with a patient whose upper...

Page 3: ...only accessories recommended by BMC Never operate the humidifier if any of the parts are damaged if it is not working properly or if the humidifier has been dropped or mishandled Do not use the humidi...

Page 4: ...indicates the maximum water level for safe operation Outlet Port Connect the flexible tubing coaxial 22 mm here Water Chamber The removable water chamber holds the water for humidification It has a s...

Page 5: ...ector Plug this connector into the power outlet on the BMC RESmart Device Set Up 1 Connect the Humidifier to the RESmart device Make sure they are connected completely 2 Place the RESmart device and h...

Page 6: ...ifier will work automatically The yellow indicator light on the humidifier will turn on Press the humidifier button can turn off or restart the humidifier 5 The ideal humidity setting depends on room...

Page 7: ...mes with clean water and allow to air dry 4 Fill the water chamber and close the silica gel cover Inspect the water chamber for any leak or damage Replace the water chamber if any damage is present Hu...

Page 8: ...condensing Atmospheric Pressure 860 to 1060 hPa 500 to 1060 hPa Power Requirements 24V DC 1 0A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Sh...

Page 9: ...the System When traveling the optional CPAP carrying case is for carry on luggage only The carrying case will not protect the humidifier if it is put through checked baggage Security Stations For eas...

Page 10: ...missions CISPR 11 Group 1 Class B The RESmart uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 11: ...for power supply lines 1 kV for input output lines Mains power quality should be that of a typical home or hospital Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2...

Page 12: ...re than is indicated in the device specification it may be necessary to position the RESmart further from sources of power frequency magnetic fields The power frequency magnetic field should be measur...

Page 13: ...nications equipment should be used no closer to any part of the RESmart including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transm...

Page 14: ...adio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sh...

Page 15: ...ive material or part BMC Medical Co Ltd will pay customary freight charges from BMC Medical Co Ltd to the dealer location only This warranty does not cover damage caused by accident misuse abuse alter...

Reviews: