InH2

®

 

Heated Humidifier

 User Manual V2.7

 

5/13 

English 

Water Chamber 

1. Turn the 

RESmart

®

 device off and allow approximately 15 minutes for the 

heater plate and water to cool. 

2. Disconnect the tubing from the water chamber. Press down on the water 
chamber and slide it out of the humidifier platform. 

3. Open the silica gel cover of water chamber and discard any remaining water. 
Fill a solution of warm water and a mild dishwashing detergent into the 
chamber, cover the silica gel cover, rock the chamber a few minutes, and then 
outflow the solution. Rinse the chamber several times with clean water and 
allow to air dry. 

4. Fill the water chamber and close the silica gel cover. Inspect the water 
chamber for any leak or damage. Replace the water chamber if any damage is 
present. 

Humidifier Platform 

1. Clean the humidifier platform by wiping with a damp cloth. Allow to air dry. 

2. Inspect the humidifier platform for any damage and replace if necessary. 

Silicone Tube of Humidifier 

Clean the Silicone Tube by wiping with an alcohol cotton stick. Allow to air dry. 

Service 

The humidifier does not require routine servicing.   
If the humidifier malfunctions, contact your home care provider immediately. 
Never attempt to open the humidifier’s enclosure. If necessary, contact your 
local authorized dealer or BMC Medical Co., Ltd. for technical support and 
documents. 

Summary of Contents for INH2

Page 1: ...English InH2 Heated Humidifier User Manual 0123...

Page 2: ...er is only used for single patient and must not be re used on another person This is to avoid the risk of cross infection The InH2 Heated Humidifier is not intended for use with a patient whose upper...

Page 3: ...only accessories recommended by BMC Never operate the humidifier if any of the parts are damaged if it is not working properly or if the humidifier has been dropped or mishandled Do not use the humidi...

Page 4: ...indicates the maximum water level for safe operation Outlet Port Connect the flexible tubing coaxial 22 mm here Water Chamber The removable water chamber holds the water for humidification It has a s...

Page 5: ...ector Plug this connector into the power outlet on the BMC RESmart Device Set Up 1 Connect the Humidifier to the RESmart device Make sure they are connected completely 2 Place the RESmart device and h...

Page 6: ...ifier will work automatically The yellow indicator light on the humidifier will turn on Press the humidifier button can turn off or restart the humidifier 5 The ideal humidity setting depends on room...

Page 7: ...mes with clean water and allow to air dry 4 Fill the water chamber and close the silica gel cover Inspect the water chamber for any leak or damage Replace the water chamber if any damage is present Hu...

Page 8: ...condensing Atmospheric Pressure 860 to 1060 hPa 500 to 1060 hPa Power Requirements 24V DC 1 0A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Sh...

Page 9: ...the System When traveling the optional CPAP carrying case is for carry on luggage only The carrying case will not protect the humidifier if it is put through checked baggage Security Stations For eas...

Page 10: ...missions CISPR 11 Group 1 Class B The RESmart uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 11: ...for power supply lines 1 kV for input output lines Mains power quality should be that of a typical home or hospital Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2...

Page 12: ...re than is indicated in the device specification it may be necessary to position the RESmart further from sources of power frequency magnetic fields The power frequency magnetic field should be measur...

Page 13: ...nications equipment should be used no closer to any part of the RESmart including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transm...

Page 14: ...adio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sh...

Page 15: ...ive material or part BMC Medical Co Ltd will pay customary freight charges from BMC Medical Co Ltd to the dealer location only This warranty does not cover damage caused by accident misuse abuse alter...

Reviews: