6
Service Manual for RESmart
®
CPAP and AutoCPAP V1.5
3.2 Hidden Function
1. Clearing Patient Data
(Available on version 1.28 and later)
Clearing the data of the RESmart CPAP and Auto-CPAP unit requires use of
a hidden function. Caution: Once the data is cleared, there is no possibility of
recovering erased data.
To clear all recorded data and restore the unit to initial factory default settings,
enter the Patient Menu by depressing ramp for 3 seconds or more. Advance
the menu using the “+” sign until you reach the date option. Manually change
the date to reflect: Year = 2097, Month = 07, Day =01, Hour = 0, Minute =26.
Press On/Off button when “Hour” setting is displayed. It is not necessary to
save the setting.
Important note: After data has been cleared, it will be necessary to
re-set the device, reviewing all patient and clinician settings (i.e. enabling
iCode, correcting to current date, etc.).