background image

 

Troubleshooting Tables 

 

If difficulties should be encountered in the course of operation, and no major instrument malfunctions can 
initially be detected, please recheck the symptoms, referring to the tables provided below, before contacting 
your service representative. 
 

1. Optical 

Symptoms Causes  Countermeasures 

Revolving nosepiece not in click-stop 

position(objective not centered optical path)

 

Revolve to click-stop position 

(swing the objective correctly into 

the optical path)

 

Field diaphragm not centered

 

Centering

 

Field diaphragm closed too far

 

Open properly

 

Darkness at the 

periphery, or uneven 

viewfield brightness.

 

Dirt or dust on the lens (condenser, 

objective, eyepiece, slide)

 

Cleaning

 

Dirt or dust on the lens (condenser, 

objective, eyepiece, slide)

 

Cleaning

 

Dirt or dust on the slide.

 

Cleaning

 

Dirt or dust in the 

viewfield 

Condenser position too low

 

Correct positioning

 

No coverglass attached to the slide 

Attach coverglass 

Coverglass too thick or thin 

Use coverglass of specified 

thickness(0.17mm) 

Slide upside-down. 

Turn over the slide 

Immersion oil on dry system objective 

(especially 40X) 

Cleaning

 

Dirt or dust on the lens (condenser, 

objective, eyepiece, slide) 

Cleaning

 

No immersion oil used on immersion system 

objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture and field diaphragm 

opened too far 

Close properly 

Dirt or dust on the entrance lens. 

Cleaning 

Condenser aperture closed too far 

Open properly 

Poor image quality 

(low resolution or 

contrast) 

Condenser position too low 

Raise to the position where the 

diaphragm image is in focus 

Revolving nosepiece not in click-stop 

position 

Revolve to click-stop position 

Image dark on one side 

Floating specimen 

Fasten securely 

Specimen rises from stage surface 

Place it stable 

Image shifts during 

focus 

Revolving nosepiece not in click-stop 

position 

Revolve to click-stop position 

Image tinged yellow 

Daylight filter not used 

Use daylight filter 

Summary of Contents for C2

Page 1: ...ce making you familiar with the instrument entirely it is strongly recommended that you read this manual carefully before using the microscope For the further reference please place this manual in a position where nearby the worktable and fetched easily BMS C2 Microscope Operation Instruction ...

Page 2: ...OVERVIEW ...

Page 3: ...he specifications of power supply which marked on the microscope base The power supply must comply with the specifications otherwise the microscope will be damaged 5 Changing the Light bulb Before replacing the light bulb with new one first turn off the light switch on the microscope base then take off the power plug and wait until the bulb cools down Only halogen bulb 6V 20W is workable on this m...

Page 4: ...ing environment Room temperature 0 to 40 Relative humidity max 85 High temperature and humidity will cause molds dew formation and damage z Avoid placing the microscope under the direct sunlight or the room light Bright circumstances deteriorate the specimen image z Avoid placing the microscope in dusty surroundings When the microscope is not in use cover it up with a vinyl cover ...

Page 5: ... the cover to their original place Eyepiece tube Mount the eyepiece tube onto the arm and tighten the clamp screw Eyepieces Insert the eyepieces into the eyepiece sleeves and tighten the clamp screw Objectives Lower the stage to its full limit then screw the objectives into the revolving nosepiece Daylight Filter Press the filter into the bottom of the auxiliary lens and then insert the auxiliary ...

Page 6: ...he right left field of view can be composed one 5 Eyepiece Diopter Adjustment Firstly observe at right tube with right eye adjust coarse fine focusing knob to image clearly Secondly observe at left tube with left eye adjust the diopter control 1 to image clearly 6 Objective Selection Rotate the revolving nosepiece to the desired objective magnification assuring it click stops into position 7 Apert...

Page 7: ... If it is opened to a larger aperture than necessary extraneous light will enter the viewfield reducing image quality and contrast 9 Oil immersion Observation When using the 100X objectives it is an oil immersion type requiring the application of immersion oil between its lens tip and the coverglass If air bubbles enter the oil layer a poor image will result To remove the air bubbles rotate the no...

Page 8: ...ing No coverglass attached to the slide Attach coverglass Coverglass too thick or thin Use coverglass of specified thickness 0 17mm Slide upside down Turn over the slide Immersion oil on dry system objective especially 40X Cleaning Dirt or dust on the lens condenser objective eyepiece slide Cleaning No immersion oil used on immersion system objective Use immersion oil Air bubbles in immersion oil ...

Page 9: ...7mm Specimen image movement unsmooth Mechanical stage not securely fastened Tighten all fasteners Binocular image not integrated Interpupillary distance not correctly adjusted Adjustment Incorrect diopter adjustment Correct adjustment Excessive eye fatigue Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Symptoms Causes Countermeasures No electrical power Chec...

Page 10: ...ute alcohol and petroleum benzine are quite inflammable Take great care when handling them and when setting the power switch on and off Be very careful with fire 2 Cleaning Painted or Plastic Surfaces z Avoid use of any organic solvents such as alcohol ether thinner etc to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent did may be clean...

Reviews: