OZONE MONITOR BMT 932 Manual, Rev. 02/2021
30
In the more complex Link-Mode it is possible to view and change many parameters interactively.
Data transmission on the serial interface are always ended with a Carriage Return (CR, dec. 13).
User-Mode
If
Timed/Polled
is set to Timed, there is an output every
Time Interval
seconds (s. configura-
tion of the RS-232 on p. 27). In case Polled is set, the BMT 932 expects the input of a question-
mark ('?', without CR), to which it responds with a block of data. User mode data output always
follows the format in the following example:
26.04.07,13:54:22,0.001ppm,0.000ppm,0.000ppm,0.000ppm,N/A,
0.000 ppm,0000
Data are separated by commas. Contents of a data block:
Date (format as set in
Date Format
)
Time as hh:mm:ss
Six Ozone concentrations including unit, depending on the setting of
Units-Ozone
(posi-
tion of decimal point depends on range). Six blocks of concentration data are always shown,
even in one or three channel instruments, starting with channel one and ending with channel
six. If there are no data available, "N/A” is displayed instead of a concentration value. This
could have the following reasons:
Channel does not exist (1 or 3 channel instruments)
Warmup
No result available yet
16-Bit status information coded hexadecimally
The 16 bits of the status info have the following meaning:
Bit
Meaning
0 (LSB) Lamp Low Warning
1
Lamp Low Error
2
Lamp Off Error
3
Utility Scrubber Error
4
Reserve Scrubber Error
5
Overpressure Error
6
Overrange Error
7
EEPROM Error
9
Warmup
10
Lamp High Error
11
Low Flow Error
12
Lamp High
14
Low Alarm
15
High Alarm
Bits 8, 12 and 13 are not used. A more detailed description of these errors is found on p. 32.