OZONE ANALYZER BMT 964 Manual, Rev. 04/2021
19
Input/Output
Simulate Analog Out
For test purposes both analog outputs can be set to their max. (10 V / 20 mA) re-
spectively min. (0 V / 4 mA) values. With the program BMT 964 Link any voltage
and current can be put out.
RS-232
This menu item is used to configure the serial interface. The User-
Baud rate
can
be set to one of the following values:
2400 Baud
4800 Baud
9600 Baud (default)
19200 Baud
38400 Baud
The setting of
Timed/Polled
decides, if the output of data on the serial interface
is done automatically in a certain time interval (Timed) or if a block of data is sent
only on request (character '?', without CR). If the operating mode is set to Timed,
a data block is sent every
Time Interval
. Minimum interval is 1 s, max. interval is
99 s. A detailed description of the serial interface can be found on page 10.
Gas Parameters
Warning:
A change of this parameter will influence measurement results!
Molecular Weight
When using the BMT 964 it is important to give the instrument information about
the carrier gas. Two different carrier gases can be chosen:
Oxygen, or oxygen from PSA, molecular weight 31.9988
Air (molecular weight 29.0)
If the customer does not specify a particular molecular weight when ordering, it
will be set to oxygen.
Time/Date
Time
Here the time of day is set in the format hh:mm:ss.
Date Format
Display of date can be switched from European (DD.MM.YY) to American
(MM/DD/YY) notation.