background image

recording  time  is  5  hours  before  you  will  need  to  recharge  your  O2  Vibe 

battery.

 

 

2.6

 

How to Set

-

up your 

O2 Vibe Device Settings

 

After 

Bluetooth

 

is 

paired,  follow  the  instructions  in  the  APP  to  set  your 

target

 

number  of  step  or 

Goal

,  switch 

ON/OFF

 

V

ibration

,  adjust  the 

vibration 

I

ntensity

 

and the lowest SpO2 

T

hreshold

 

(default is 90%) to set up 

your 

O2 Vibe

 

Wrist Pulse Oximeter.

 

2.7

 

Data Storage

 

Your  O2  Vibe  stores  4  recordings,  Sleep  or  Fitness,  so  it  is  recommended 

you  synch  your  O2  Vibe  device  to  your 

O2  Vibe  mobile  APP

 

at  minimum 

every  other  day.    T

he  O2  Vibe  will  delete  the  oldest

 

recording  when  you 

record 

your  5

th

 

recording.    All  recordings  on  your  O2  Vibe  device  will 

automatically  sync  and  transfer  to  your 

O2  Vibe  mobile  APP

 

when  you 

connect via Bluetooth.

 

3.

 

Using O2 Vibe Sleep & Fitness Monitor 

3.1

 

Prior to Use

 

Before  using 

O2  Vibe

see 

Warnings  and 

C

aution

s

 

on  page  1,  and 

p

ay 

attention to the following points in order to obtain precise measurements

.

 

Ÿ

 

The finger inserted in 

SpO

2

 

sensor must be clean to ensure proper 

reading.

 

Ÿ

 

Any  of  the  following  conditions  may  cause  inaccurate 

measurements, including but not 

limited to:

 

o

 

Flickering or very bright 

fluorescent 

light

ing

;

 

o

 

Poor blood circulation;

 

o

 

Low hemoglobin;

 

o

 

Hypotension,  severe  vasoconstriction,  severe  anemia  or 

hypothermia;

 

o

 

Any  tests  recently  performed  on  you  that  required  an 

injection of intravascular dyes.

 

Ÿ

 

Th

e O

2 Vibe

 

may not work if you have poor circulation. Rub your 

finger to increase circulation, 

reposition

 

your finger or

 

move the 

sensor  to  another 

finger.

   

Your  thumb  is  best,  then  your  index 

finger on down to your smallest finger.

 

Ÿ

 

The 

O2  Vibe

 

measures  oxy

gen  saturation  of  functional 

hemoglobin. High levels of dysfunctional hemoglobin (caused by 

sickle  cell  anemia,  carbon  monoxide,  etc.)  could  affect  the 

Summary of Contents for O2 Vibe

Page 1: ...ntroduction and Intended Use 1 2 Getting Started 4 3 Using O2 Vibe Sleep Fitness Monitor 6 4 O2 Vibe Care and Maintenance 10 5 Troubleshooting 12 6 Limited Warranty 13 7 Specifications 13 User s Manual Tracking Your Oxygenation in a Smart Way TM ...

Page 2: ...gs and Cautions Do not use this device during MRI examination The O2 Vibe has no alarm do not use when alarms are required Please DO NOT squeeze the sensor or apply excessive force to the sensor or cable as shown below to ensure longer use and accuracy of recordings Never submerge the device in water or other liquids Do not clean the device with acetone or other volatile solutions Consult your doc...

Page 3: ...oisture in the sensor o improperly applied sensor o incorrect sensor type o poor pulse quality o venous pulsations o anemia or low hemoglobin concentrations o carboxyhemoglobin o methemoglobin o dysfunctional hemoglobin 1 2 How O2 Vibe Sleep Fitness Monitor Works The O2 Vibe Sleep and Fitness Monitor O2 Vibe measures oxygen saturation pulse rate steps and motion The O2 Vibe works by shining two be...

Page 4: ...f solid foreign objects 12 5mm diameter Against dripping 15 tilted Follow operating instructions Caution SpO2 Percent oxygen saturation Heart rate Sleep mode Heart rate Fitness mode Sleep mode Fitness mode Bluetooth connected to O2 Vibe APP Insert the ring sensor plug into the device Slide your thumb or finger into the ring sensor SN Serial number found on the underside of the device No alarm syst...

Page 5: ...ds to power OFF the device You change screens and modes as well using this button by pressing quickly for new menus or functions Note To reset your O2 Vibe device please hold ON OFF button down for 8 seconds continuously 2 3 Charging Caution The O2 Vibe device cannot be used for any measurement during charging Please fully charge the battery for three hours before first use Fully discharging the b...

Page 6: ... O2 Vibe device has a built in Bluetooth radio that connects with your O2 Vibe mobile APP to store Sleep or Fitness recordings Update device software display exercise metrics via mobile Dashboard and set SpO2 Threshold Vibration On Off and Intensity using your O2 Vibe mobile APP when O2 Vibe device is connected via Bluetooth In Sleep Mode you do not need to connect to O2 Vibe mobile APP via Blueto...

Page 7: ... Fitness Monitor 3 1 Prior to Use Before using O2 Vibe see Warnings and Cautions on page 1 and pay attention to the following points in order to obtain precise measurements The finger inserted in SpO2 sensor must be clean to ensure proper reading Any of the following conditions may cause inaccurate measurements including but not limited to o Flickering or very bright fluorescent lighting o Poor bl...

Page 8: ...e O2 Vibe screen display and O2 Vibe mobile APP if connected via Bluetooth Real time display uses the O2 Vibe mobile APP Dashboard function 3 If you don t see your readings the signal strength may be low reposition or switch to another finger try again 4 Remember do not squeeze your ring sensor forcibly as shown here as it may damage sensor 3 3 Turning ON and Using your O2 Vibe for First Time To t...

Page 9: ...plug into the device in 10 seconds it will stop measurement automatically and show your O2 score and how many times Drops your SpO2 was lower than Threshold you set After 2 minutes the O2 Vibe will automatically turn OFF In Fitness Mode if the sensor ring plug is disconnected from your device or the sensor ring slips from your finger you will see the O2 device begin a countdown from 10 seconds and...

Page 10: ... will be in Normal Mode where your screen shows SpO2 and Heart Rate measurements To change your mode press the button or if you wait approximately 2 minutes the O2 Vibe screen will go dark and into Hibernation Mode automatically to eliminate bright light while you are sleeping While in Hibernation Mode you may press the button to change the screen display between measurements and Sleep Mode screen...

Page 11: ... turning OFF your O2 Vibe device or remove the thumb ring sensor for more than 10 seconds You may also review a detailed report of your results on the O2 mobile APP after you sync your recording to your Apple or Android smartphone iPad or tablet via Bluetooth see Pairing Bluetooth on page 5 4 O2 Vibe Care and Maintenance With proper care maintenance and following usage instructions the expected us...

Page 12: ...cations o locations where device is exposed to direct sunlight o locations with high temperatures or levels of moisture o locations in combustible environments or heavy contamination o locations near sources of water or fire or o locations that are subject to strong electromagnetic influences ...

Page 13: ... APP within 2 minutes after powering ON or after measurement is complete Make sure your O2 APP is within reasonable range of your O2 Vibe device SpO2 or heart rate does not display Sensor positioned incorrectly Reposition sensor on your thumb or try another finger starting with Index finger The thumb is recommended Finger is cold Warm finger by rubbing Finger is wet Dry finger and sensor Excessive...

Page 14: ...cations EC Directive MDD 93 42 EEC R TTE 1999 5 EC ROHS 2 0 2011 65 EU Degree protection against electrical shock Type BF Environmental Item Operating Storage Temperature 5 to 40 C 25 to 70 C Relative humidity noncondensing 10 to 95 10 to 95 Barometric 700 to 1060 hPa 700 to 1060 hPa Degree of dust water resistance IP22 Physical Weight Less than 50 g main unit Display 0 69 OLED Connector Micro D c...

Page 15: ...79 3 PR range 30 to 250 bpm PR accuracy 2 bpm or 2 whichever is greater Sleep Recording Recording time Up to 10 hours Data storage SpO2 pulse rate Measurement summary O2 score Total duration 90 STAT Average saturation Lowest saturation Summary No abnormal detected blood oxygen drop detected Unable to analyze Pedometer Range 0 to 99999 steps Vibrator Event Steps low oxygenation Threshold Oxygenatio...

Page 16: ...255 00917 00 Version A Contents of this manual are subject to changes without prior notice Copyright 2016 Shenzhen Viatom Technology Co Ltd All rights reserved O2 Vibe Sleep Fitness Monitor Knowing and Tracking Your Health Tracking Your Oxygenation in a Smart Way ...