35
Immunity
test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment-
guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2.
±8kV direct
& indirect
contact;
±15kV air
discharge.
±8kV direct
& indirect
contact;
±15kV air
discharge.
Floors should be
wood, concrete
or ceramic tile. If
floor are covered
with synthetic
material, the
relative humidity
should be at
least 30%.
Electrical fast
transient/
burst IEC
61000-4-4.
±2 kV for
power
supply lines.
Not
applicable.
Not applicable
(For INTERNALLY
POWERED ME
EQUIPMENT.
Surge IEC
61000-4-5.
± 1 kV
line(s) to
line(s).
Not
applicable.
Not applicable
(For INTERNALLY
POWERED ME
EQUIPMENT.
Guidance and manufacturer’s declaration -
electromagnetic immunity
The device is intended for use in the electromagnetic
environment specified below. The customer or the user
of the device should assure that it is ensure in such
environment.