37
Immunity
test
IEC
60601
Test
level
Comp-
liance
level
Electromagnetic
environment-guidance
Radiated
RF IEC
61000-
4-3.
10V/m
& table
9.
10V/m
& table
9.
Portable and mobile RF
communications equipment
should not be used close to any
part of a blood pressure monitor
Monitor, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of
the transmitter. Recommended
separation distance.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to
the transmitter manufacturer
and d is the recommended
separation distance in metres
(m). Field strengths from fixed
RF transmitters, as determined
by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency
range
b
Interference may occur in
the vicinity of equipment marked
with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from
structures, objects and people.