2. Important Information
- This non-contact fever thermometer should only be used for the purposes described in this
instruction manual
- This non-contact fever thermometer is adjusted and calibrated in the production, so there
is no need of any additional adjustment, regulation or similar
- According to many clinical research studies, without complying with specific regulations,
there is no chance to compare the measurement results made on different types of
thermometers and on different parts of human body
- This thermometer can be used in an ambient temperature range b 10 to 40 degree
Celsius only
- Do not expose this thermometer to the electric shocks and to extreme temperature
conditions of higher than 50 degree Celsius or lower than 0 degree Celsius
- Do not store or use this thermometer on relative humidity higher than 85 %
- With this fever thermometer you can make the body temperature measurement on
following body parts: behind the ear lobe, inside the ear, on the temple, orally, rectally,
axiliary, forehead
- After inserting a new batteries it is necessary to wait approximately 10 minutes prior to first
measurement
- Time distance between each following measurement must be at least 30 seconds!
- Do not use this thermometer near large electromagnetic fields such as mobile phones and
similar
- Keep this thermometer away from water or any liquid, away from heat or direct sunlight
- Do not drop or knock this thermometer and do not use it damaged
- Measurement accuracy can be affected by many factors ( for instance when the forehead is
covered by hair, perspiration, cap or scarf )
- Keep the measuring distance 1 – 10 centimeters only!
- This thermometer should be left in the room 15 – 20 minutes before measurement
Protective glass in front of the thermometer must be kept clean at all times !
NEVER focus this thermometer towards the eyes!
- Select “BODY” mode to measure the body temperature
- Select “SURFACE” mode to measure the surface temperature
- In case of any problems or uncertainties do not open this thermometer, please do not
hesitate to contact the authorized service center, of which address could be found in this
instruction manual.
3. Features
- Precise non-contact measurement of human body temperature, but also surface (certain
surfaces) temperatures
- Selectable BODY or SURFACE temperature
- Automatic memorization of last 32 measurement results
- Set Alarm adjustment
- Automatic switch off
- Selectable Celsius or Fahrenheit temperature range
- Display resolution 0,1 degree
Backlight LCD display
4