16
2
)
Fast Mode:
Key Description:
Notice
:
1) When the operation must be kept hand piece and skin close contact, in order to
achieve better results.
2) The energy output is controlled by two levels, that is, the "Ready" button is
pressed, and then the energy output.
3) At the end of the treatment, click the "Standby" button, so that the device is in
standby mode.
4) Setting-up energy output must be gradually increased, and the body feels
comfortable, avoid high energy, otherwise it may lead to skin damage. When it’s in
operation, please maintain communication with patients, according to patients’ feeling to
adjust parameters at any time.
Note: once the patient's skin mild burns, must immediately stop operating, and in
time to carry out the necessary treatment to prevent wound infection.
5) Before the operation the headpiece must be disinfected, can use 75% medical
alcohol to wipe, then dry, prohibited the use of sharp scratch head surface, working head
prohibited bump, avoid damage.
Prevent the hand with water, causing permanent damage to the laser.
In addition
Touch these buttons to change handpiece
Touch these buttons to change Mode of Big handpiece
Touch these buttons to change Skin & Hair Color
Touch these buttons to Adjust Energy ,Pulse and Frequency