The device may be cleaned by gently wiping with a damp cloth pre-moistened with mild
soap and water. Never immerse the device in water or other liquids.
Wipe lead wires with a damp cloth if they become soiled.
To properly store the device for an extended period of time, remove the battery from the unit.
Put the unit and accessories in the carrying case and store in a cool, dry location.
Cleaning the Device
Troubleshooting
If the device does not function properly:
1. Make sure the battery is properly installed or replace the battery. Be sure to observe
proper polarity markings when replacing the battery. If the indicator light is not on
when the unit is turned on, replace the battery and recheck.
2. If the intensity has been adjusted and there is no stimulation, check that the lead
wires are properly connected and the electrodes are in place. If the unit appears to
be functioning and no stimulation occurs, the lead wires or electrodes may need to
be replaced.
3. If the battery appears to be charged and the unit is not functioning, turn both
Amplitude Control knobs to the OFF position (counterclockwise) for about 7 seconds.
Next, gradually turn the Amplitude Control knobs clockwise until stimulation is felt.
If device is still not working, turn the unit off and contact BodyMed
®
.
If any other problems occur, please consult or return the device to BodyMed
®
.
Do not repair a defective device.