8. TROUBLESHOOTING
Should any malfunction occur while using the device, check
whether the parameters are set appropriately for therapy, and
adjust the control correctly. Please see the following table:
Malfunction Common Reasons
Countermeasure
No display
after
replacing
the battery
1. There’s a foreign body in the
battery compartment. 2. The
battery has been used up or
installed incorrectly. 3. There
is foreign body in the battery
interface. 4. The battery is not
the right model or something
goes wrong with the battery
interface 5. Exception reset.
1. Check and clean the
compartment 2. Replace the
new battery or install the battery
correctly. 3. Check and clean the
interface 4. Replace the battery
with the right model.
No
sensation
of
stimulation
1. The electrode does not
connect well to the skin.
2. The electrode does not
connect well to the stimulator.
3. The battery is used up.
4. The skin is too dry.
1. Check and re-paste it on skin.
2. Check the connection. 3.
Replace the battery. 4. Wipe the
electrode and the skin with a wet
cotton cloth.
Automatic
halt in the
treatment
1. The electrode loses
connection with the skin.
2. If the battery is used up.
1
. Check and place the electrode
properly on the skin. 2. Replace
the battery.
Rash or
tickle on
the skin
occurs
in the
treatment
1.The treatment time lasts
too long.
2.The electrode does not stick
well to the skin.
3. The interface of the
electrodes is dirty or dry.
4. The skin is sensitive to the
electrode.
1. Do the treatment once a day
and shorten the treatment time.
2. Check and stick the electrode
well. 3. Wipe the electrode with
a wet cotton cloth before use.
4. Check your allergic history.
Please change the sticking place
or shorten the treatment time. If
your skin is over-sensitive, you
should stop the treatment or go
to see a doctor.
7.4 Cleaning the Electrodes Cords:
Clean the electrode cords by wiping them with damp cloth. Coating
them lightly with talcum powder will reduce tangles and prolong
the life.