37
Test Specifications for ENCLOSURE PORT IMMUNITY to
RF Wireless Communications Equipment (Table 9)
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b
) Maximum
Power (W)
Dis-
tance
(m)
Immunity
Test Level
(V/m)
385
380-
390
TETRA 400
Pulse
modulation
b
)
18Hz
1.8
0.3
27
450
430-
470
GMRS 460,
FRS 460
FM c)±5kHz
deviation
1kHz sine
2
0.3
28
710
704-
787
LTE Band 13, 17
Pulse
modulation
b
)
217Hz
0.2
0.3
9
745
780
810
800-
960
GSM800/900,
TETRA 800, iDEN
820, CDMA 850,
LTE Band 5
Pulse
modulation
b
)
18Hz
2
0.3
28
870
930
1720
1700-
1990
GSM1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1,3,
4,25; UMTS
Pulse
modulation
b
)
217Hz
2
0.3
28
1845
1970
2450
2400-
2570
Bluetooth, WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b
)
217Hz
2
0.3
28
5240
5100-
5800
WLAN 802.11
a/n
Pulse
modulation
b
)
217Hz
0.2
0.3
9
5500
5785
NOTE: If it is necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The
1 m test distance is permitted by IEC 61000-4-3.
a. For some services, only the uplink frequencies are included.
b. The carrier shall be modulated using a 50 % duty cycle square wave signal.
c. As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used
because it does not represents actual modulation. It would be worst case.