background image

Warnings and Precautions

WARNINGS
•  The device must be kept out of reach of children.
•  The safety of the device for use during pregnancy or delivery has not been 

established.

•  Do not place electrodes on front of the throat. This may result in spasms of the 

laryngeal and pharyngeal muscles.

•  Do not place the electrodes over the carotid nerve (the front and sides of the neck).
•  The device is not effective for headaches.
•  Caution should be used when applying the device to patients suspected of having 

heart disease. Further clinical data is needed to show if there are adverse side 
effects on those with heart disease.

•  The device may interfere with electronic monitoring equipment (such as ECG 

monitors and ECG alarms).

•  Electrodes should not be placed over the eyes, in the mouth, or internally.
•  These devices have no curative value. The Interferential is a symptomatic treatment 

that suppresses the sensation of pain which would otherwise serve as a protective 
mechanism.

•  Interferential devices should be used only under the continued supervision of a 

physician or licensed practitioner.

•  Do not use on broken skin.
•  Effectiveness is highly dependent upon patient selection by a person qualified in pain 

management.

•  If the device treatment becomes ineffective or unpleasant, stimulation should be 

discontinued until re-evaluation by a physician or licensed practitioner.

•  Always turn the device off before applying or removing electrodes.
•  The device does not have AAP/APG protection. An explosion hazard is possible if 

used in the presence of explosives, flammable materials or flammable anesthetics.

ADVERSE REACTIONS
•  Skin irritation and electrode burns are potential adverse reactions. Stimulation 

should be stopped and electrodes removed until the cause of the irritation or burns 
can be determined. 

•  Isolated cases of skin irritation may occur at the site of electrode placement 

following long-term application. Always properly clean skin before use. If skin 
irritation occurs, discontinue use. Do not resume use of the ZZAIF400 unit until you 
have seen your physician or licensed practitioner.

Summary of Contents for ZZAIF400

Page 1: ...ENGLISH READ THIS INSTRUCTION MANUAL CAREFULLY BEFORE USE Instruction Manual Interferential Current Stimulator Reorder Number ZZAIF400...

Page 2: ...y 3 Indications and Contraindications 4 Warnings Precautions 5 About the Device 6 Device Controls 6 Attaching the Lead Wires 7 Electrode Selection and Care 7 Connecting the Device 8 Cleaning the Devic...

Page 3: ...ctrical stimulation General Description Your BodyMed ZZAIF400 unit is warranted to be free from defects in materials and workmanship occurring within one year from date of purchase when used in strict...

Page 4: ...nt to flow trans cerebrally through the head Indications and Contraindications Limited Product Warranty continued BodyMed MAKES NO OTHER WARRANTY EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION ANY IM...

Page 5: ...protective mechanism Interferential devices should be used only under the continued supervision of a physician or licensed practitioner Do not use on broken skin Effectiveness is highly dependent upo...

Page 6: ...for Mode Time Width and Rate Press the front of the cover and pull down to open the cover Amplitude Controls The Amplitude Control Knobs are located on the top of the unit under the cap The Amplitude...

Page 7: ...to maintain stimulation and prevent skin irritation The packaging will provide instructions for care maintenance and proper storage of the electrodes Be sure to use the electrodes provided by BodyMed...

Page 8: ...ad wire into the channel output receptacle to be used pushing plug in as far as it will go 5 Select treatment settings Check that the unit is set to the proper settings as recommended by your physicia...

Page 9: ...the unit This will reveal the battery compartment 2 Remove the old battery from the device 3 Place a new battery in the compartment Note the proper polarity alignment indicated on the battery and the...

Page 10: ...able 0 80mA peak into 500 load each channel constant current Carrier Frequency 4000 Hz fixed CH1 Modulating Frequency 4004 4160 Hz Adjustable CH2 IF Frequency Mode Constant Mode 4 160 bps Adjustable A...

Page 11: ...lse frequency varies from 30 to 60 of the set pulse frequency Six seconds at the lower frequency and six seconds at the higher frequency The transition is abrupt square wave function 3 6 6 ramped shif...

Page 12: ...Distributed by BodyMed Hudson Ohio 44236 1 866 528 2152 Made in Taiwan Reorder Number ZZAIF400...

Reviews: