15
Calibration/Span Adjustment
Features
Calibration/adjustment can be
performed only when:
– there is no load on the balance/scale,
– the scale is zeroed, and
– the internal signal is stable.
If these conditions are not met, an error
message is displayed.
The weight required for calibration/span
adjustment is displayed.
Standard calibration weights for selected
models: to remove, see instructions on
page 7.
Press (F) to select a different weight
value.
To cancel the procedure: press and
hold the (ENTER) key (
≥
2 sec.).
Example:
Calibrate/adjust span of the balance/scale (here: model BEB61)
Step
Key (or instruction)
Display
1.
Switch on the balance/scale
(ON/OFF)
2.
Zero the balance/scale
(ZERO)
3.
Start calibration
(CAL)
The preset calibration weight
is displayed without the weigh unit
(in this example, 5000 g)
4.
To select a different calibration
weight value
(F) repeatedly
5.
Start calibration/span adjustment
(ENTER)
After the zero point is stored, the required
calibration weight flashes on the display.
6.
Place the required calibration
weight on the balance/scale
The readout stops flashing if the weight
is applied within the defined time limit and
tolerance. If the weight value is accepted,
the display stops flashing and the stability
symbol
appears on the display.
7.
Remove the calibration weight
8.
Calibration/span adjustment has
been completed
Calibration is recommended after initial installation and each time the
balance/scale is moved.