1 Step : Minimum Division Set
2 Step : Maximum Capacity Set
3 Step : Rated Output value Set
4 Step : Zero Calibration
5 Step : END
GhYd%
- Function :
A]b]aia8]j]g]cbGYh
Range → 1~ 100
A step to set up a division value.
"d" means "Division" and "xx" means a division capable of displaying.
Also this value will be displayed as 1-2-5-10-20-50-100 by each key.
A step to set up a Decimal point is Function mode.
So, it will be go to the next step recording the position.
Key
Display
Description
“SAVE” key :
Increase
“ENTER” key :
Store and move
Into next menu
Sim.CAL. condition.
1 kg (Decimal point : 0)
0.01kg (Decimal point: 2)
☞ REF 1.The minimum division means the value of one division.
☞
REF
2. External resolution is obtained by division the min. division by the
maximum capacity. Set the resolution to be within 1/30,000.