4. OPERATION ANOMALIES
In case of abnormal operation and before the intervention of the aftersales
service,
Findings
Causes, Checkings and solutions
Impossible to switch on the appliance with
the On/Off switch.
Electric supply cut.
-
Check the connection of the supply
cable
- Check the fuses and/or the circuit-
breaker located before the power plug.
Complete stop of the appliance during a
cycle.
Power cut of the area
-
Check the fuses and/or the circuit-
breaker located before the power plug.
The compressor operates intermittently.
Compressor security thermostat.
-
Check the operation of the
condenser
fan
.
-
Check the condenser is not clogged
.
-
Check the ambient temperature
.
Ventilate if necessary.
The group unit (compressor and
condenser fan) operates intermittently.
Security pressure sensitive switch cut.
- Check the operation of the condenser
fan
.
-
Check the condenser is not clogged
.
-
Check the ambient temperature
.
Ventilate if necessary.
The cycle duration is superior to the
required time.
Loading too important.
-
Reduce the loading
.
High ambient temperature.
-
Ventilate.
Frozen evaporator.
-
Defrost.
Evaporators fans blocked.
-
Clean them.
Condenser too clogged.
-
Make it clean