Instructions n°648261 –
Page 10
3.3.3 TABLE OF PARAMETERS
Code
Parameters
Mini
Maxi
Usine
(2)
PA
Password to access level 2
(set on –19)
-55
99
0
/
= PROBES
I
1
Core probe calibration
(correction by 1/8°C)
-55
99
0
/3
Number of core probe points
0 = 1 point, 1 = 5 points
0
1
1
/5
Ambient probe calibration
(correction by 1/8°C)
0
6
0
/8
Temperature scale
0 = °F, 1 = °C
0
1
1
o
= MACHINE CONFIGURATION
o0
Cycle type
0 = Chill, 1 = Chil/fr
0
1
1
o1
Door contact
0 = No, 1 = Yes
0
1
1
o2
Choice of cycle when closing the door
0 = No, 1 = Yes
0
1
0
o3
Automatic prefunction cycle when powered up
0 = No, 1 = Yes
0
1
0
r
and
c
= REGULATOR CONFIGURATION
r0
Prefunction cycle regulation differential
(°C)
1
15
3
r1
Prefunction cycle set-point
(°C)
-55
99
-15
c0
Preservation after cycle regulation differential
(°C)
1
15
3
(1)
c1
Chilling cycle duration without core probe
(mn)
30
180
120
(1)
c2
End of chilling cycle set-point with core probe
(°C)
0
99
10
c3
Chilling cycle preservation temperature set-point
(°C)
-55
99
3
(1)
c4
Chilling/freezing cycle duration without core probe
(mn)
30
300
270
(1)
c5
End of chilling/fr. cycle set-point with core probe
(°C)
-55
0
-18
c6
Chilling/freezing cycle preservation temperature set-point
(°C)
-55
99
-20
c7
Ambient / Core probe insertion check
(with cb = 1)
(°C)
0
99
5
(1)
c8
Temperature level for countdown start
(°C)
0
99
63
c9
End of cycle buzzer alarm duration
(s)
0
99
3
(2)
cA
Core probe temperature
(read only)
--
--
--
cb
Core probe insertion test
(according to c7 and cE)
0 = No, 1 = Yes
0
1
1
(2)
cc
Fragile products ambient probe set-point
(°C)
-55
99
-15
cd
Temperature level for weak/strong ventilation cycle
(°C)
-55
99
12
cE
Core probe insertion test duration
(with cb = 1)
(s)
1
99
30
C
= COMPRESSOR PROTECTIONS
C0
Compressor delay at power up
(mn)
0
15
0
C1
Min. time between 2 starts of the compressor
(mn)
0
15
0
C2
Min. break after stop of the compressor
(mn)
0
15
0
C8
0
99
85
d
= DEFROST
d0
Time between 2 defrosts
(h)
0
99
8
d1
Defrost type
0 = Ventil, 1 = Elect
0
1
0
d2
End of defrost temperature (ambient probe)
(°C)
0
30
5
d3
Max. defrost duration (safety time)
(mn)
1
99
30
d4
Defrost at the beginning of the chilling cycle
0 = No, 1 = yes
0
1
0
d5
Defrost delay at the beginning of the pres. cycle
0 = Without (mn)
0
99
0
(2)
d6
Defrost at the beginning of the prefunction cycle
0 = No, 1 = Yes
0
1
1
d7
Draining time
(mn)
0
99
0
db
Time basis
0 = H mn, 1 = mn s
0
1
0
(1)
: Parameters corresponding to legislation (H-L indication if the cycle doesn’t comply with these settings).
(2)
: Configuration parameters on level 1, accessible to user.