GasFinderFCr Operation Manual — Installation
July 17th 2017
Condensation
Condensation on the surface of the window is caused by the window temperature
falling below the dew point. This may occur in areas of high humidity or large
temperatures differences between day and night. The enclosures are well
sealed but condensation can sometimes occur on the inside of the window.
Condensation can usually be controlled by the following measures:
Fasten a cover or rain hood over the enclosure to shield the
•
window from the effect of radiation cooling at night. This also
prevents rain from striking the window.
Use a desiccant pack inside the enclosure. This will only prevent
•
condensation on the inside.
Place a heater in the enclosure. The heater can be 12W to 50W
•
depending on the size of the enclosure and the conditions of
operation and is controlled by a thermostat to prevent overheating.
The heater raises the temperature of the enclosure and prevents
condensation forming either on the inside or the outside of the
window. To minimise heat loss the enclosure is insulated. This
option requires power to be supplied and there are heaters available
for use in hazardous areas.
SS Enclosures with
Heaters and Rain Hoods
4 retros
4 retros
heater
heater
thermostat
thermostat
4 retros and heater in
insulated enclosure
Radiation Shield
Summary of Contents for GasFinder FC
Page 50: ...7 8 Boreal Laser Inc GasFinderFCr Operation Manual Troubleshooting July 17th 2017...
Page 56: ...9 2 Boreal Laser Inc GasFinderFCr Operation Manual Appendices July 17th 2017...
Page 71: ......
Page 73: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm...
Page 75: ......
Page 77: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......