8
RANGE
ACCURACY
RES.
2 V
± 0,8% of rdg ± 3 dgt
100 μV
20 V
1m V
200 V
10m V
750 V
± 1,2% of rdg ± 3 dgt
1 V
Input resistance 10 MΩ Frequency range 40-1000 Hz.
Overload protection AC 750 V Eff ective value or DC 1000 V
peak value (constantly in all ranges).
Display: Corrected centre point corresponds to the eff ective
value in a sinusoidal form.)
HINT:
• In case the voltage range is unknown, you must set the
FUNCTION switch in the highest measurement range
and then reduce the range gradually if required.
• If the fi gure «I» now appears on the display, the meas-
urand lies outside the range set. The FUNCTION switch
must be switched to a higher range.
• The maximum input voltage is 750 VAC eff ective value.
Higher voltages cannot be measured.
• Proceed extremely cautiously with the measurement of
high voltages.
Measuring direct current (ADC)
• Connect the BLACK lead to the COM connector socket.
Connect the RED lead to the «VΩmA» connector socket.
For measurements between 200mA and 10 A, remove
RED lead to «10 A» connector socket.
• Set the FUNCTION switch on the desired «A» setting and
connect the leads in series to the power circuit to be
measured. Observe correct polarity (red is + and black is
-), otherwise the display shows a minus sign before the
value. The technical current direction of the RED lead to
the BLACK lead is displayed at the same time with the
current strength.
• Read off the value in (milli-) Ampere.
RANGE
ACCURACY
RES.
200 μA
± 0,8% of rdg ± 1 dgt
0,1 μA
2 mA
1 μA
20 mA
10 μA
200 mA
± 1,2% of rdg ± 1 dgt
100 μA
10 A
± 2% of rdg ± 1 dgt
10 mA
Overload protection 0.5 A/250-V cut-out, 10 A-range not
protected.
HINT:
• In case the current range is previously unknown, you
must set the FUNCTION switch in the highest measure-
ment range and then reduce the measurement range
gradually if required.
• If the fi gure «I» now appears on the display, the measur-
and lies outside the set measurement range. The FUNC-
TION switch must be switched to a higher measurement
range.
• The 10 A range is not protected with a fuse. Do not,
therefore, measure for longer than 10 seconds.
Measuringthe resistance (Ω Ohm)
• Connect the BLACK lead to the «COM» connector socket
and the RED lead to the «VΩmA» connector socket.
(Hint: The polarity of the red lead is then «+».)
• Setthe FUNCTION switch on the desired «Ω» measuring
range.
RANGE
ACCURACY
RES.
200
Ω
± 0,8% of rdg ± 3 dgt
0,1
μ
A
200
Ω
± 0,8% of rdg ± 3 dgt
0,1
Ω
2 K
Ω
± 0,8% of rdg ± 1 dgt
1
Ω
20 K
Ω
10
Ω
200 K
Ω
100
Ω
2 M
Ω
1 K
Ω
Overload protection 250 V DC or AC Eff ective value, for maxi-
mum 15 seconds
• Connect the measuring tips to the components to be
measured. Ensure that the component is not linked to
the other components. Do not touch the measuring tips
so as not to aff ect resistance values.
• Read off the measured value in Ω (Ohm).
• While measuring the resistances, current from the in-
ternal battery is used. This current consumption diff ers
according to the set measurement range.
HINT:
• In case the measurand exceeds the highest value in the
set measurement range, the display shows «I». Select
a higher measurement range. If the resistance is about
1 MΩ and higher, the measuring device can take a cou-
ple of seconds to stabilise. This is normal in the measure-
ment of high resistances.
• If the input is not connected, say in a broken power
circuit, the display shows the fi gure «I». This means that
the measurement falls outside the range.
• In case the resistance to be measured is connected to
a power circuit, you must switch off the voltage before
beginning the measurement and see that all capacitors
are discharged.
Diodes measurement (Conducting-state voltage)
• Connect the BLACK lead to the «COM» connector socket
and the RED lead to the «VΩmA» connector socket.
(Hint: The polarity of the red lead is then «+».)
• Setthe FUNCTION switch on the
-range and connect
the measuring tips to the diode to be measured. In the
measurement of diodes or transistors, the polarity of the
measuring tips determines whether the forward direc-
tion or the blocking direction is measured. The displayed
value is the conducting-state voltage drop.
- The tripping current is 0.8 mA.
- In case the measuring tips are not or wrongly con-
nected to the diode i.e. in the blocking direction, the
display shows the fi gure I.
Summary of Contents for 91271150
Page 1: ...BMM 800 91271150 RU FR DE GB Digital Tester 4 Digital Multimeter 7 Multim tre num rique 10 13...
Page 2: ...3 1 0 1000 V 0 750 V 0 10 A 0 15 kg 0 2 M 2...
Page 16: ...17 RU FR DE GB Garantiebedingungen Warranty terms Conditions de garantie...
Page 22: ...23 BORT GLOBAL LIMITED BORT GLOBAL LIMITED 2 Defort 5 SDS BORT GLOBAL LIMITED GmbH 21 10719...
Page 26: ...30...
Page 27: ...31...