Area of use and safety instructions
The boso TM-2450 blood pressure measuring device operates according
to the oscillometric principle of measurement. The device is used to take
readings over a 24-hour period. It is suitable for use both in hospitals
and in community medicine. It is not suitable for use on young children,
infants or on unconscious patients without supervision.
N.B.
Do not allow the air tube to be compressed or its diameter to be
reduced.
Excessively frequent readings can impair circulation and
consequently lead to injury.
The cuff must not be fitted over wounds as this could lead to
further injury.
Ensure that the cuff is not placed on an arm of which the arteries
or veins are undergoing or have undergone medical treatment,
e.g. a shunt.
In the case of women who have had a mastectomy, do not place
the cuff on the arm on the side on which the operation took place.
Medical devices being used on the same arm at the same time
may malfunction.
The device is not protected against the possible effects of high-
frequency (HF) surgical devices.