Click to open
measurement
channels in data
view
Double-click
to open
online view
Click to open
calibration
window
Analog and physical
value
5. Click on ‘Calibrate adjustment points’ to open calibration window.
6. Apply the desired physical condition to the sensor (e.g. by applying a force
on the wheel).
7. Enter the physical value in the value column of the desired calibration point
(e.g. 745 N).
8. Press the ‘Calibrate’ button of the desired calibration point.
9. Repeat for all curve points.
10. Click ‘Close’ when done.
The calibration curve is displayed in the online view.