10. Error Messages
Measured
temperature
too low
Ambient
temperature
too high
Ambient
temperature
too low
Error function
display
Blank display
Empty battery
indication
Displays „L“ when
measured
temperature lower
than 0 °C or 32.0
°F.
Displays „H“ in
combination with
the „
L
“ when am-
bient temperature
is higher than 40.0
°C or 104.0 °F.
Display „L“ in
combination with
the „
M
“ when
ambient tempera-
ture is lower than
10 °C or 50 °F.
When item has
malfunction. If mal-
function occurs fre-
quently, please con-
tact manufacturer
(see page 31)
Please check if the
battery has been
inserted corecctly.
Also check poliari-
ty (<+> and <–>) of
battery.
If the battery icon
is shown on the
display, the batter-
ies should be
replaced immedi-
ately.
Display / Problem
Display
Meaning
Measured
Temperature
too high
Possible cause
and fault
remedy
Displays „H“ when
measured
temperature higher
than 100.0 °C or
199.9 °F.