Clinician Remote Control Directions for Use
Clinician Remote Control Directions for Use
90930936-07 REV A 22 of 125
en
Reset Stimulator
You can use the RC to reset the Stimulator.
From the Clinician Menu:
1. Select
Reset Stimulator. The Reset Stimulator? screen displays.
2. Select
Yes to reset the Stimulator.
or
Select
No to cancel the action.
Exit Clinician Mode
From the Clinician Menu:
1. Select
Exit Clinician Mode. The System Settings screen displays.
Troubleshooting
Telemetry Help
The RC has an effective communication range of up to 91.4 cm (36 inches) with the Precision Spectra
Stimulator and a nominal effective communication range of 55.9cm (22 inches) with the Precision Novi
Stimulator. Common sources of interference, such as televisions and computer monitors, can affect
wireless communication. You will continue to have access to the RC menus even if the RC is outside
its effective communication range. If you enter a command to change your settings or programs while
the remote control is outside of its effective communication range, you will have several seconds to
move the RC within range of the Stimulator to send the command. After the command is received
by the Stimulator, the remote control will make a double-beep then display the requested change on
the screen. If the RC is unable to communicate with the Stimulator, the Telemetry Error message will
appear: