10 Technical specifications
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
101
Monopolar Cutting – MetraLOOP
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Cutting
MetraLOOP"
Effect
P (W)
1
100
2
150
3
200
Table of power output P [W] as a function
of the setting "Monopolar Cutting
MetraLOOP"
Rated load resistance= 500
Effect
U (Vp)
1
650
2
650
3
650
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar