10 Technical specifications
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
107
Monopolar Cutting – GastroCut Knife medium
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Cutting
GastroCut Knife medium"
coag phase
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Cutting
GastroCut Knife medium"
cut phase
Effect
P (W)
coag phase
P (W)
cut phase
1
3
300
2
5
300
3
9
300
4
14
300
5
17
300
6
22
300
7
26
300
8
30
300
9
40
300
Table of power output P [W] as a
function of the setting "Monopolar
Cutting GastroCut Knife medium"
Rated load resistance= 500
Effect
U (Vp)
1
650
2
650
3
650
4
650
5
650
6
650
7
650
8
650
9
650
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Cutting GastroCut Knife medium"
(idle mode)