10 Technical specifications
116
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Monopolar Coagulation – Argon flex. pulse
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Coagulation
Argon flex. pulse"
= 40 W
Measurement of resistances
Diagram of power output P [W] as a
function of the load resistance R [
] for
the setting "Monopolar Coagulation
Argon flex. pulse"
= 80 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Coagulation Argon flex. pulse"
Rated load resistance = 500
Effect
U (Vp)
1
4400
2
4400
3
4400
Table HF output voltage U [Vp] as a
function of the setting "Monopolar
Coagulation Argon flex. pulse" (idle
mode)
The puls frequency changes with the effect setting. The higher the effect
level, the faster the pulse sequence.
Effect 1: 1 Hz, effect 2: 5 Hz, effect 3: 10 Hz
The mode "Argon flexible" is paused due to the pulse sequence.