5 Operation
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
31
5.4.
Configuring output currents
All selection windows are closed after 30 seconds without assumption of
the modification.
If selection screens are present, buttons outside the screen are inactive.
In addition, activations not possible in this case.
A change to the currently loaded program, e.g. by adjusting the output, is
indicated by the information “changed” below the program name.
5.4.1. Selecting the mode
Figure 5-7: Modes monopolar cutting
1. To select the type of current for monopolar cutting, first select one of the two
socket connectors on the left side.
2. Press the yellow icon under the "Mode" button.
A selection screen appears for the available modes, and the rim of the
associated connector starts blinking.
3. Select the desired mode by pressing the corresponding button.
- or -
Deactivate the mode by pressing the button “mode off”.
4. Press the "?" button for more information on this selection.
5. Additional selection options are available in the selection window. To select
indication-related modes, press the “Indications” button.
6. Confirm your selection by pressing the "OK" button.
The main screen will be displayed.
- or -