5 Operation
46
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
5.6.
Monopolar cutting modes
5.6.1. Standard
In this mode a high-performance HF current with a low crest factor is used for
cutting biological tissue. ARC CONTROL quickly adjusts the power output to the
minimum required level in response to variations in tissue type and changes in the
cutting area or speed.
Application areas
Cutting tissue with low electrical resistance, such as muscle tissue or vascular
tissue.
Cutting or preparing fine structures
Suitable instruments
Needle electrodes
Knife electrodes
Spatula electrodes
Sling electrodes
5.6.2. Micro
This mode is used for electrosurgical cutting using micro-electrodes. It enables
extremely fine control of the power level and precise work.
Application areas
Pediatric surgery, neurosurgery, plastic surgery
Suitable instruments
Micro needle electrodes