6 Detecting and correcting faults
84
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
6.2.
Fault indications for EASY monitoring
Fault indications are displayed in three stages (green, yellow and red) when
problems occur.
When working with a split neutral electrode, the following faults may occur:
EASY
monitoring
Cause
Indication
Corrective measures
Flashes yellow Significant increase in
resistance
Depending on the
indication, there may be
heating under the neutral
electrode
–
Stopping the application is
not necessary.
Check the proper
application of the neutral
electrode.
Switches from
green to
continuous red
A significant problem
occurred when the
monopolar current was
activated
An acoustic
signal sounds.
A warning
message
appears on the
display
Check the neutral electrode
and neutral electrode cable
(see Section
electrode monitoring
(EASY monitoring)
, page
Check the neutral electrode
cable for proper connection
and external damage.
Loosened electrode
An acoustic
signal sounds.
A warning
message
appears on the
display
Reattach the neutral
electrode. If the fault
persists, replace the neutral
electrode.