8 Maintenance and repair
86
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
8. Maintenance and repair
8.1.
Maintenance
DANGER
Infection hazard!
Carry out a surface disinfection and wrap the device in addition to
the shipping packaging material before allowing the device to leave
the hospital or office to avoid spreading germs and infections.
Check the device, the device trolley and the accessories (e.g. foot switch,
cable) after each use for damage or defects. In particular, make sure that the
insulation is intact on all cables.
Do not use any damaged device, damaged device trolley or damaged
accessories.
Replace defective accessories immediately.
Have the safety inspection for the device performed once a year. Please
consult and comply with the respective service instructions for additional
technical information.
8.1.1.Safety inspection
Safety inspections must be performed once a year.
The next safety inspection date of ARC 400 can be displayed in the dialog,
see “BOWA Service” dialog, page 69
A warning message appears during system start-up if a safety inspection is
due.
Press OK to confirm this message.
Any shorter safety inspection cycles specified in national regulations
must be observed.
The device and accessories may be inspected only by persons who have the
required training, knowledge or experience and who can perform the
inspection independently.
With regard to the safety inspection, you must comply with the country-
specific rules and regulations.
The inspector documents the inspection results and measured values
corresponding to the printed inspection protocol.
In the case of severe deviations from the values of the attached final acceptance
report, or if the specified maximum values were exceeded:
Send the HF device to the service center, see section Technical service,
page 88.