background image

Ultrasonic Cleaner

Maintenance Schedule

* Always observe universal precautions when there 
might be potential exposure to infectious bio-burden.

Recommended

n

     Do perform recommended maintenance to ensure proper performance and 

longevity of your ultrasonic cleaner.

n

     Do pre-rinse instruments to remove larger bio-burden particles. This is kind 

of like flossing teeth to remove larger particles so the brushing can focus on 
removing the smaller particles!

n

     Do download and print our Instrument cleaning instructions guide found at 

www.brandmax.us/msds-literature.

n

     Do use our Burr Basket (part # “BB1”) to hold small items including burrs.

n

     Do operate ultrasonic unit with liquid at the Max Fill line level.

Not Recommended

(

will

 

void

 

your

 

warranty

)

DON’Ts

n

     Do not remove the filter (except during 

recommended maintenance).

n

     Do not pick up the unit and drain, unit 

must remain stable at all times.

n

     Do not put any items directly in the tank, 

all items must be placed inside basket.

n

     Caution! - Using the enzyme tablets 

and/or granular enzymes can damage 
solenoid and void your warranty!  We 
recommend using a liquid enzymatic 
cleaner in your ultrasonic cleaner.

DOs

TRI-CLEAN

BrandMax

®

DO THIS...

DAILY

WEEKLY

MONTHLY

Change ultrasonic solution when visibly soiled 
or once a day, which ever is more often.

P

   Rinse the tank thoroughly and wipe down with 
soft cloth at the end of each work day.

P

Daily, flush drain tube with hot water to ensure 
smooth draining cycles. 

P

Clean inside of tank with mild detergent once a 
week. Be sure to rinse thoroughly so detergent is 
not transferred to cleaning cycle.

P

Perform Foil Test 

Please contact BrandMax for testing details.

P

Replace filter every 3 to 6 months or when 
slightest signs of damage or wear appear. 

Part  # “FIL1” or “FIL2” 
Please specify model number and serial 
number when ordering.

P

Reviews: