Ultrasonic Cleaner
Maintenance Schedule
* Always observe universal precautions when there
might be potential exposure to infectious bio-burden.
Recommended
n
Do perform recommended maintenance to ensure proper performance and
longevity of your ultrasonic cleaner.
n
Do pre-rinse instruments to remove larger bio-burden particles. This is kind
of like flossing teeth to remove larger particles so the brushing can focus on
removing the smaller particles!
n
Do download and print our Instrument cleaning instructions guide found at
www.brandmax.us/msds-literature.
n
Do use our Burr Basket (part # “BB1”) to hold small items including burrs.
n
Do operate ultrasonic unit with liquid at the Max Fill line level.
Not Recommended
(
will
void
your
warranty
)
DON’Ts
n
Do not remove the filter (except during
recommended maintenance).
n
Do not pick up the unit and drain, unit
must remain stable at all times.
n
Do not put any items directly in the tank,
all items must be placed inside basket.
n
Caution! - Using the enzyme tablets
and/or granular enzymes can damage
solenoid and void your warranty! We
recommend using a liquid enzymatic
cleaner in your ultrasonic cleaner.
DOs
TRI-CLEAN
™
BrandMax
®
DO THIS...
DAILY
WEEKLY
MONTHLY
Change ultrasonic solution when visibly soiled
or once a day, which ever is more often.
P
Rinse the tank thoroughly and wipe down with
soft cloth at the end of each work day.
P
Daily, flush drain tube with hot water to ensure
smooth draining cycles.
P
Clean inside of tank with mild detergent once a
week. Be sure to rinse thoroughly so detergent is
not transferred to cleaning cycle.
P
Perform Foil Test
Please contact BrandMax for testing details.
P
Replace filter every 3 to 6 months or when
slightest signs of damage or wear appear.
Part # “FIL1” or “FIL2”
Please specify model number and serial
number when ordering.
P