Chapter 4: Installation and Setup
Input Power Requirements
4-8
100-214-252 Rev. A
4.5 Input Power Requirements
The input power requirements for the SLPt Cell Disruptor are:
•
117 VAC, 50/60 Hz (North American model)
•
200-245 VAC, 50/60 Hz (Export models only)
The SLPt Cell Disruptor is equipped with an IEC-type power cord connector. The unit requires a single-
phase, three-wire, 50/60 Hz power source.
WARNING
!
To prevent the possibility of electrical shock, always plug the SLPt Cell Disruptor unit into a
grounded power source.
The system is fuse-protected with a replaceable glass fuse, 5x20mm, slow-blow type (refer to the data tag
on the system. This fuse should never blow under normal operating conditions. The fuse holder is found on
the rear of the unit, as part of the IEC power connector.
4.6 Electrical Connections to Equipment
All of the connections to the SLPt Cell Disruptor are made to the rear of the unit using industry-standard
connectors. Refer to
Figure 2.3, Back Panel of SLPt Cell Disruptor,
on page 2-5 for connector locations.
4.6.1 Power Cord
North American units are shipped with a 3-conductor 117 Volt cordset (NEMA 5-15P to IEC jack). It con-
nects to an IEC-type connector on the rear of the unit. The plug end connects to your main voltage recepta-
cle, which should be properly fused (depending on your site requirements). It requires a conventional
NEMA 5-15R receptacle for installation.
Export units are shipped with a standard Harmonized European cordset (having an IEC-type jack).
CAUTION
!
If your cordset does not match your main power receptacle, verify that you have the correct
voltage available. Do not connect the system if the voltage rating of the unit is incorrect for
your location, as this can damage the unit.
DOC EXPIRES 12PM 9/16/2007. Article or Material must comply with the requirements
stipulated by RoHS in its current version of 1/27/03