SLPt Cell Disruptor
Chapter 6: Operation
User Manual
System Modes
100-214-252 Rev. A
6-5
Pulse Start (PS)
To set the method of trigger switch duration, select Off (0) if you desire to hold the trigger signal throughout
the duration of the process cycle. If the start signal is lost, the cycle is aborted and no alarm is produced.
Select On (1) if you wish to start the cycle with a pulse (minimum duration of 10ms). After sending the
pulse, the sonics will remain on for the duration of the cycle. Pressing again will abort the current process
cycle.
NOTE
i
i
In both modes, the start signal must be removed before the next cycle.
Auto Reset (Ar)
To set the mode for resetting all alarms, select Off (0) to reset using the front panel Reset key or through the
rear mounted external user I/O connector J2, utilizing pins 1 and 7.
When On (1) is selected, a reset signal is not required. All functions of the power supply are accessible.
Panel Trigger (PT)
Refer to the following Table to set the method of starting a cycle for any mode of operation,
6.3.3 Control Menu
After power up, the display will go to the ready state and displays rdY (ready).
There are two groups of addressable functions, Mode and Set.
The current Mode LED will be on. Pressing the Mode button will sequence to the next mode in a continuous
loop. The selected mode’s LED will come on and the corresponding Set (parameters) associated with the
mode will light.
Table 6.1
Panel Trigger (Pt) Off (0), Pulse Start (PS)
Off (0)
Start signal must come from User Input J2, and must be
held throughout the cycle. When signal stops, cycle ends.
Panel Trigger (Pt) On (1), Pulse Start (PS)
On (1)
Start signal can come from either the front panel Mem-
brane Keypad or the User Input J2. When a signal is
applied (20 ms min) and released the cycle will continue
until the key is pressed again to Stop.
Panel Trigger (Pt) Off (0), Pulse Start (PS)
On (1)
Start signal must come from User Input J2. When a signal
is applied (20 ms min) and released the cycle will con-
tinue until J2 is triggered again to stop.
Panel Trigger (Pt) On (1), Pulse Start (PS)
Off (0)
Start signal can come from either the MEmbrane Keypad
Start/Stop key or J2, and must be held throughout the
cycle. When key is released, cycle ends.