SLPt Cell Disruptor
Chapter 6: Operation
User Manual
Operational Sequence
100-214-252 Rev. A
6-7
6.4 Operational Sequence
NOTE
i
i
While Configuration Menu parameters can be specified each time the unit is powered up, the
default factory settings are correct for most applications, and are enabled at power up. If there
is a need to re-configure, press and hold the ENTER key (6) while powering up.
To configure the SLPt:
1. Press and hold Enter key while turning on power.
2. When in the Configuration Menu, use the Left (8) arrow key to navigate to either Pulse Start, Panel Trig-
ger or Auto Reset, as required.
3. Use the Up (9) arrow key to toggle between Off (0), and On (1), and apply to selected item as desired.
4. Press ENTER to accept choices and exit the Configuration Menu.
Set Power Supply Wattage
1. Press TEST (5) while powering up. All Set LED’s (3) will blink.
2. Set wattage to match power supply wattage, 150 for SLPt.
3. Press ENTER to go to ready mode.
6.4.1 SLPt in Continuous Mode
Process Sequence
1. Turn On power, wait for display (1) to show Ready (rdY).
2. Select Mode (2) > Continuous (2A). Amplitude LED (3D) will light.
3. Set Amplitude % from 10 to 100% using arrow keys.
4. Press ENTER.
5. Send Start pulse by shorting pins 2 and 6 of external user I/O, J2 to start cycle, or by pressing the Start/
Stop key, as per configuration.
NOTE
i
i
entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters.