Chapter 7: Maintenance
Troubleshooting Charts
7-4
100-214-252 Rev. A
Table 7.1
System Trouble Analysis Chart
Symptom
Probable Cause
Corrective Action
Main power fuse fails or
circuit breaker trips when
system is plugged into
electrical outlet.
• Cordset has failed
• Replace cordset
• Power Switch has failed.
• Line filter has failed.
• Return for repair
Display screen does not
appear when unit is
turned on. Fan does not
operate.
• System is unplugged or main
power is missing
• Correct power problem
• Unit fuse has blown (it
should not under normal
conditions)
• Replace fuse
• Cordset has failed
• Replace cordset
• Power Switch has failed
• Line filter has failed
• Unit failed due to connection
to incorrect input voltage
• Return for repair
Fan does not operate
when system is turned on.
Display screen appears.
• Fan motor has failed
• Return for repair
Fuse fails when system is
turned on.
• Fuse is under-rated
• Mains Voltage is incorrect
• Fan motor has failed
• Power supply module has
failed
• Unit failed due to connection
to incorrect input voltage
• Verify voltage source is
correct. Damage may
occur if connected to
wrong voltage source
• Replace fuse with cor-
rect value and retry; or
return for repair
Ultrasonic power is not
delivered to the horn.
• Ultrasonic power supply
module has failed
• Digital controls have failed
• Return for repair
Unusual noise from tool
when ultrasonics are on
• Tool is loose, or it is contact-
ing a solid object
• Reposition tool
• Remove, examine, and
clean tool; reinstall tool
• Tool has failed
• Replace tool