Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
►
Set aside the product if it is damaged.
8.
Maintenance
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Detection
Cause
Remedy
Charger not working
“Power On” indicator not illumi-
nated
Mains power cord not plugged in
Plug mains power cord into the
device connector
7
on the charger
and into the socket of the mains
power supply.
Power cord faulty
Replace power cord.
Fuses blown
Replace fuses, see Replacing fuses.
Charger defect
Replace product.
Defective indicator
During the switch-on check, not all
indicator elements are illuminated
("Power On" indicator and indicator
fields for charging bays )
Charger defect
Replace product.
Battery does not charge
Battery inserted, charging progress
indicator not illuminated
Charger contacts dirty
Clean the contacts, see Validated
reprocessing procedure.
Battery defective
Repeat the charging process in
another charging bay. If the prob-
lem occurs again, replace product.
Charger defect
The charging process does not start
up either with a different battery.
Replace product.
Battery inserted, “Action request”
indicator illuminated
Excessive battery temperature
detected during charging
Remove the battery from the
charging bay, let it cool down and
restart charging. If the problem
occurs again, replace battery.
Battery defective
Repeat the charging process in
another charging bay. If the prob-
lem occurs again, replace battery.
Charger defect
The charging process does not start
up either with a different battery.
Replace product.
Summary of Contents for Aesculap GA320
Page 2: ......
Page 3: ...9 5 6 7 4 2 3 1 8...
Page 88: ...86 Aesculap Acculan 4 Acculan 4 GA320 6 6 1 7 3 2 3 1 2 3 Acculan 4 GA320 7 6 2...
Page 89: ...87 GA346 GA666 GA676 A B 1 1 7 7 1 7 2 VAH FDA CE 7 3 7 A B...
Page 90: ...88 Aesculap Acculan 4 Acculan 4 GA320 7 4 7 5 I 1 T C F t I 1 Meliseptol Wipes sensitive...
Page 91: ...89 7 6 8 B Braun Aesculap 9 Aesculap 7...
Page 139: ...137 GA346 GA666 GA676 A B 1 1 7 7 1 CJK CJK 7 2 VAH FDA CE 7 3 7 A B...
Page 140: ...138 Aesculap Acculan 4 Acculan 4 GA320 7 4 7 5 RT 1 1 C F I 1 Meliseptol Wipes...
Page 141: ...139 7 6 8 B Braun Aesculap 9 Aesculap ON 7 ON...