RF
cable
Instruction
1. Set up the structure according to the figure 1.
2. Signal generator setup:
signal type:1. 3G 3Carriers
3FA, PAR
:
8.0dB
2.
ats_3gpp_fdd_fwd_tm1_64ch_sc0_v5pt1
3. Spectrum analyzer setup:
RBW
→
30KHz, VBW
→
300KHz, Center frequency: 1960MHz, SPAN
→
40MHz, Sweep
time: 1S Spectrum Analyzer’s offset value according the attenuation from PA output to
input port of spectrum analyzer.
4. Power meter must to be calibration and set the offset value according to attenuation
from PA output to power sensor.
5. Input PA voltage setup +30V.
6. Turn on the signal generator.
7. Than confirm standard ACLR peculiarity.