18
Breathing Circuits
Always use a 22mm bore breathing circuit when the High Flow alarm is set
above 80 lpm.
Always use a 15mm bore breathing circuit when the High Flow alarm is set
below 80 lpm.
The type of breathing circuit used will depend on the mode of ventilation selected:-
Breathing circuits to be assembled in the order A-B-C-D from the ventilator end.
In
CPAP
,
Pressure Support
and
Pressure Control
modes, a single limb circuit with
an exhalation port will be required.
DO NOT USE THIS TYPE OF CIRCUIT FOR
IPPV MODE
22mm circuit, Part Number 0792
Breathing circuit volume = 570 ml including filter.
---------------------------------------------------------
In
IPPV
mode a single limb circuit with an exhalation valve will be required. This
type of circuit will have a 22mm diameter main tube and a small tube to connect the
exhalation valve to the EVC (exhale valve control) port on the Nippy .
THIS
IS THE ONLY TYPE OF CIRCUIT RECCOMMENDED FOR USE IN IPPV MODE
22mm circuit, Part Number 0793
Breathing circuit volume = 600 ml including filter.
---------------------------------------------------------