45
NIPPY+ Battery run time test record
Use this table to record the battery run times when performing the monthly battery
health test.
Replace the battery when the run time is less than 75% of a new battery or the
battery is 2 Years old.
Fully charge the battery, connect the breathing circuit and using your normal settings
run the ventilator until the low battery alarm operates. Record the run time in the
table.
NIPPY
Serial Number:…………………………..
Date
Runtime
– Internal
battery
Runtime
– External Battery
If applicable