7
Operating the STAP
operating your device.
• Be sure to always ground yourself before handling the unit
• Be sure to always connect the unit to earth ground with the provided grounding
cable
• Be sure the ground wire between the detector cover and the main board main-
tains good contact
7.2 Changing the Filters
Follow the procedure below to change the two filters inside the STAP.
Figure 10:
First step to
changing STAP filters.
• Shut off the STAP control software and disconnect power.
• Remove the two 4-40 bolts show in Fig. 10 that pass through the detector board
and hold the two main STAP body pieces to each other. The allen wrenches re-
quired to change the filters are in the ship kit.
• Pull the detector board and underlying aluminum STAP body piece away from
the unit to reveal the two stainless steel filter sealing pieces shown in Fig. 11.
The two 10 mm diameter filters are held underneath the two stainless steel
pieces.
83-00036-01A
STAP 9406