background image

1300 276 642

www.bromicrefrigeration.com.au

7

Power Supply

•  The appliance must not be plugged into a multiple outlet power board.
•  The appliance is supplied with a flexible power cord fitted with a 3-pin plug.
•  The socket used must be properly earthed and fuse protected.
•  See the rating label inside the cabinet for supply voltage, frequency and current draw.
•  Power cable is rated 70°C. It is recommended that the cable should not come into contact 

with parts over 50°C.

•  Do not turn o

ff

 power before consulting the person responsible.

•  Mark the power source clearly, so the refrigerator is not unplugged or turned o

ff

 

accidentally.

Before Operating

•  If the appliance has been sharply tilted or transported in the horizontal position, it must 

be left to stand in the upright position and not plugged in for 1 hour.

•  Remove all the packaging material before using the appliance.
•  Start the appliance only if it has been installed as described.

1300 276 642

www.bromicrefrigeration.com.au

6

NOTE: 

The maximum ambient temperature for the refrigerator is 30°C. Conditions 

outside this temperature will result in cabinet temperatures outside the required 2 

to 8°C range. Bromic recommends that the refrigerator is installed at an ambient 

temperature of 20 to 25°C.

4. INSTALLATION

Locating the Cabinet

Ensure the following instructions are observed as this is vital for proper and trouble-free 

operation.
•  Do not install the cabinet outdoors.
•  Locate the appliance in a dry and well ventilated place.
•  The unit must be positioned on a level surface.
•  Do not install in the vicinity of heating radiators or other heat sources.
•  Do not install the appliance where it is exposed to direct sunlight.
•  Do not install the appliance where it is exposed to warm draughts or open doors.
•  Keep the air intake and outlets free of obstructions.
•  Adequate allowance should be made for door openings.

Ventilation

Ensure the following instructions are observed as this is vital for proper and trouble-free 

operation.
•  Allow 100mm clearance between walls and the appliance.
•  Ensure that the condenser fan is free to take in ambient air from the room.
•  Adherence to minimum clearances is necessary

SIDE VIEW

100mm

100mm

100mm

100mm

FRONT VIEW

Summary of Contents for Medifridge MED0140GD

Page 1: ...23 05 2016 V2...

Page 2: ...otification This Installation operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration 1 PREFIX 3 2 IMPORTANT NOTES WARNINGS 4 3 PRODUCT OVER...

Page 3: ...ction manual Any safety information will be accompanied by the following safety alert symbols DANGER WARNING IMPORTANT READ THIS MANUAL CAREFULLY before installing or servicing this product Improper i...

Page 4: ...ith reduced physical sensory or mental capabilities or lack of experience and knowledge unless they have been given supervision or instruction concerning use of the appliance by a person responsible o...

Page 5: ...m Reversible No Shelves No of Shelves 4 Style of Shelves Wire open rung 7mm apart Depth of Shelf 3 Shelves at 365 mm 1 Shelf at 235 mm Adjustable Yes Power Supply Voltage 240V Frequency 50Hz Input Pow...

Page 6: ...l ventilated place The unit must be positioned on a level surface Do not install in the vicinity of heating radiators or other heat sources Do not install the appliance where it is exposed to direct s...

Page 7: ...ble is rated 70 C It is recommended that the cable should not come into contact with parts over 50 C Do not turn off power before consulting the person responsible Mark the power source clearly so the...

Page 8: ...frigeration circuit is operating During the start up delay 3 minutes the LED is flashing The LED indicates that the refrigeration circuit is defrosting Setting the temperature NOTE The controller is f...

Page 9: ...Type Corrective Action HI Cabinet high temperature alarm Check if there is adequate ventilation Check if there is free circulation of air inside the cabinet Check and clean condenser coil and fan Che...

Page 10: ...l advise is given Some products are sensitive to ultraviolet UV light please see storage guidelines from the manufacturer before removing from packaging Do not store food and other goods in the refrig...

Page 11: ...e the products and transfer to a suitable location prior to cleaning the inside of the cabinet Do not use any aggressive or abrasive cleaning agents Never use a water hose or a high pressure jet to cl...

Page 12: ...bient temperature too high B Inadequate ventilation C Wall clearances below minimum distance D Dirty condenser coil E Ice build up on evaporator coil A Relocate unit B Keep ventilation slots free C Ad...

Page 13: ...1300 276 642 www bromicrefrigeration com au 13 9 SCHEMATIC DIAGRAM...

Page 14: ...ions must be observed In the interest of preventing environmental nuisance and promoting re use of the high quality materials employed used commercial refrigeration equipment must be collected for exp...

Page 15: ...orage of any stock Failure to follow instructions may result in loss of or damage to stock Please refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine storage...

Page 16: ...ales 1300 276 642 Int 61 2 9748 3900 F 61 2 9748 4289 E refrigeration bromic com au NSW VIC QLD TAS WA SA NT ACT NZ BROMIC COM AU 10 Phiney Place Ingleburn NSW 2565 Australia T AU 1300 276 642 F 61 13...

Reviews: