Bronkhorst®
Instruction Manual EL-FLOW® Prestige
9.17.084N
32
Changing network settings
·
By briefly pressing the switch 5 times with intervals of up to 1 second in normal operation mode, the instrument enters a
state in which the node address and baud rate can be changed (non-Ethernet based protocols only; for Ethernet based
protocols (EtherCAT®, PROFINET), network parameters are configured by the fieldbus master and cannot be set on the
instrument).
·
Changing network parameters with the multifunctional switch is done in 3 steps, each represented by a LED indication
pattern (see table below).
·
At the start of each step, the according LED(s) start(s) flashing fast (0.1 second on, 0.1 second off). By pressing and holding
the switch, the associated action is started and the flashing slows (05. seconds on, 0.5 seconds off).
Step Pattern
Maximum
flash
count
Action
1.
Green
•
•
12
set tens of node address
2.
Red
•
•
9
set units of node address
3.
Green and red
(simultaneous)
•
•
10*
set baud rate index (number of flashes)
*) maximum count depends on the supported baud rates of the fieldbus. See the baud rate table above for supported baud
rates and associated indexes.
To execute a step, follow these instructions:
·
Press and hold the switch (flashing slows)
·
To select value 0 (zero), release the switch within 1 second, otherwise:
·
Count the number of LED flashes
·
Release the switch as soon as the required value is reached
·
In case you lose count, keep the switch pressed and wait until the flash count reaches its maximum and restarts
On completion of a step, the instrument automatically advances to the next step. When all required steps have been
completed, the instrument returns to its normal operation mode.
If the switch is not pressed within 60 seconds after starting a step, all changes in the previous steps are cancelled and the
instrument returns to its normal operation mode.
3.9
Basic parameters and properties
This section describes the most common parameters for digital communication with the instrument.
Digital parameters are most easily accessible via FlowPlot or FlowView software or a Bronkhorst® readout and control unit
(E-8000 or BRIGHT).
3.9.1
General
This section describes the most commonly used parameters for digital operation of the EL-FLOW® Prestige. Descriptions are
grouped by category in tables as shown below:
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
[type]
RW
[x]…[y]
[DDE par]
[Pro]/[Par]
[address]/[register]
In this manual, parameter names are printed in italics (reverted to
normal
where embedded in italics, like in this tip).
Type
Unsigned char
1 byte unsigned integer (0…255)
Unsigned int
2 byte unsigned integer, MSB first (0…65535)
Unsigned long
4 byte unsigned integer, MSB first (0…4294967295)
Float
4 byte floating point, IEEE 32-bit single precision, MSB first
Unsigned char [x] x byte array (text string)