background image

© Bronkhorst High-Tech B.V.

3

2

Check test-pressure

The tested pressure is stated on the instrument with a red-colored sticker.
Before installation, make sure that the test pressure is in accordance with
normal safety factors for your application. If there is no Pressure Testing
Sticker on the device or if the test pressure is incorrect, the instrument
should 

not

 be mounted in the process line and be returned to the factory.

NOTE! Tested pressure is higher than the (normal) operating pressure.

Mass Flow Meter

Mass Flow Controller

3

Check if system piping is clean

For reliable measurement always make sure the fluid stream is clean. 
Use filters to assure a particle free liquid stream or a moisture- and 
oil-free gas stream. Recommended pore size: 0,5...25 µm. If back flow 
can occur, a downstream filter and a check valve are recommended too. 
For high flow rates select a suitable filter size, to avoid too high 
pressure drop or cavitation.

Warning!

During the manufacturing process, the instruments have been tested with water. Despite the fact that the
instruments have been purged thoroughly afterwards, we cannot guarantee that the delivered instruments are
absolutely free from water droplets. Bronkhorst strongly recommends performing an additional, adequate drying
procedure for those applications where remaining water particles may cause undesired reactions such as corrosion.

4.1

Mount/install instrument properly

Install the 

mini CORI-FLOW™ ML120

 Meter/Controller in the line and tighten the fittings according to the

instructions of the supplier of the fittings. 

During operation avoid external vibrations and shocks.

4.2

Flow direction

Install the 

mini CORI-FLOW™ ML120

 in accordance with the direction of 

the FLOW arrow. The arrow for flow direction is indicated on the 

mini CORI-FLOW™ ML120

, between process fittings. 

4.3

Base mounting

Mount the 

mini CORI-FLOW™ ML120

 instrument, with screws in the body, to a rigid, stiff base body or heavy mass,

such as a wall, heavy rig or stable construction. This is essential to achieve optimal accuracy with the 

mini CORI-

FLOW™ ML120

 instrument.

By default the 

mini CORI-FLOW™ M12, M13, M14

 and 

ML-120

 will be delivered on a special mounting block for

achieving optimal accuracy.This mounting block has a mass and stiffness precisely tuned for the specific model.

Removing the mounting block will cause inaccuracy unless the instrument is firmly mounted on a vibration free, stiff
and rugged surface. Mounting on plates of machine cabinets is not recommended. 

4.4

Mounting position general

For gas and liquid 

mini CORI-FLOW™ ML120

 meters can be mounted in any position for a proper measurement.

Summary of Contents for mini CORI-FLOW ML120

Page 1: ...Doc nr 9 17 093C Date 13 05 2016 Quick Installation Guide ...

Page 2: ...nterface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 132 FLOW BUS Interface 9 17 024 9 16 133 PROFIBUS DP Interface 9 17 025 9 16 134 DeviceNet Interface 9 17 026 9 16 135 Modbus interface 9 17 035 9 16 136 EtherCAT interface 9 17 063 9 16 137 PROFINET 9 17 095 9 16 146 i www These documents can be downloaded from the website http www b...

Page 3: ...ecommends performing an additional adequate drying procedure for those applications where remaining water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW ML120 Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings During operation avoid external vibrations and sh...

Page 4: ...or leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Avoid condensation due to cold liquids gases at high humidity environments Liquid tight is not the same as gas tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using external valves which might l...

Page 5: ...l the Air gas in the tubes for liquids or to remove all possible condensation drops for gases Purging during start up In order to remove condensation drops during start up flushing with dry gas for some minutes with high flow rate while tapping the valve is recommended Special control mode for purging In case of purging of a mini CORI FLOW ML120 controller give setpoint 100 to control the valve or...

Page 6: ... and the green LED will blink fast The procedure will take approx 45 seconds 45 sec When the signal is not stable the procedure will take longer max 6 retries will be performed The red LED will blink when signal is unstable noisy 10 5 Ready When the indication is showing 0 signal and the green indication LED is burning continuously again then the zeroing action was successful Measurement check For...

Reviews: