background image

© Bronkhorst High-Tech B.V.

5

7.2

BUS/digital operation

For this procedure see description 
for RS232 operation or specific 
(optional) fieldbus. 
RS232 connection cable 7.03.366 
enables to use (free downloadable) 
tooling programs for Windows.

Example 3 

Example 4 

8

Multi-functional switch operation

Using the 2 colored LEDs and the switch on the 

mini CORI-FLOW™ ML120

,

several actions can be monitored and started. The green 

 LED is used for

status indication. The red 

 LED is used for errors/warnings. The switch can be

used to start several actions, such as auto-zero, restore factory settings and bus-
initialization actions. See specific zero-procedure part in this manual or see
'General instructions for 

mini CORI-FLOW™ ML120

' for more details.

Important warning!

 When red 

 LED blinks shortly: measuring signal is unstably or noisy: instrument is possibly

exposed to vibrations or pulsating flow.

The micro switch on top of the 

mini CORI-FLOW™ ML120

 can be operated with a thin, metal or hard plastic pin. For

example the end of a paperclip.

9

Purging

In systems for use with corrosive or reactive fluids, purging with an inert gas is absolutely necessary before use. After
use with corrosive or reactive fluids, complete purging is also required before exposing the system to air. Purge the 

mini CORI-FLOW™ ML120

 with actual fluid to get rid of all the Air/gas in the tubes (for liquids) or to remove all

possible condensation drops (for gases).

Purging during start-up

In order to remove condensation drops during start-up, flushing with dry gas for some minutes with high flow rate
(while tapping the valve) is recommended.

Special control mode for purging

In case of purging of a 

mini CORI-FLOW™ ML120

 controller give setpoint = 100 % to control the valve or the pump.

It is also possible to use special control mode = 8 to fully open the valve or set the pump at max. rpm, using a digital
interface. This will bypass the PID-controller and might be usefull when having the 

mini CORI-FLOW™ ML120

 set

to a low capacity. It will ensure you to get the highest possible flow for purging.

Warm-up time

Let the 

mini CORI-FLOW™ ML120

 warm-up for at least 30 minutes for best accuracy.

10

Zeroing

Before first use, when process conditions change significantly (especially temperature) or when the instrument has
been re-mounted (e.g. after servicing) it is recommended to perform an automatic zero action with the 

mini CORI-

FLOW™ ML120

. This action can be started manually (as described below) or via a digital interface (see document

Operation instructions digital instruments: 9.17.023). Under normal (constant) conditions it will not be necessary to
zero before each application start-up.

10.1

Set process conditions

After warming-up, pressuring and purging the system (including the 

mini CORI-FLOW™ ML120

)

 

prepare the

instrument for actual process conditions.

During zeroing avoid external vibrations and shocks.

Summary of Contents for mini CORI-FLOW ML120

Page 1: ...Doc nr 9 17 093C Date 13 05 2016 Quick Installation Guide ...

Page 2: ...nterface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 132 FLOW BUS Interface 9 17 024 9 16 133 PROFIBUS DP Interface 9 17 025 9 16 134 DeviceNet Interface 9 17 026 9 16 135 Modbus interface 9 17 035 9 16 136 EtherCAT interface 9 17 063 9 16 137 PROFINET 9 17 095 9 16 146 i www These documents can be downloaded from the website http www b...

Page 3: ...ecommends performing an additional adequate drying procedure for those applications where remaining water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW ML120 Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings During operation avoid external vibrations and sh...

Page 4: ...or leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Avoid condensation due to cold liquids gases at high humidity environments Liquid tight is not the same as gas tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using external valves which might l...

Page 5: ...l the Air gas in the tubes for liquids or to remove all possible condensation drops for gases Purging during start up In order to remove condensation drops during start up flushing with dry gas for some minutes with high flow rate while tapping the valve is recommended Special control mode for purging In case of purging of a mini CORI FLOW ML120 controller give setpoint 100 to control the valve or...

Page 6: ... and the green LED will blink fast The procedure will take approx 45 seconds 45 sec When the signal is not stable the procedure will take longer max 6 retries will be performed The red LED will blink when signal is unstable noisy 10 5 Ready When the indication is showing 0 signal and the green indication LED is burning continuously again then the zeroing action was successful Measurement check For...

Reviews: