background image

10.2

Stop flow

Make sure there is no flow through the 

mini CORI-FLOW™ ML120

 by

closing a shut-off valve in front of the instrument. At least one (shut-off) valve
in front of the 

mini CORI-FLOW™ ML120

 is required. High quality shut-off

valves are recommended for proper zero point calibration.

10.3

Start Auto-Zero

Press micro switch and hold it. After a short time the 

 red LED will go ON

and OFF, then the 

 green LED will go ON. At that moment (which is 8...12

seconds after pressing) release the switch.

10.4

Zeroing procedure

The zeroing procedure will start at that moment and the 

 green LED will

blink fast. The procedure will take approx. 45 seconds.

45 sec.

When the signal is not stable the procedure will take longer (max. 6 retries will be performed). The red 

 LED will

blink when signal is unstable/noisy. 

10.5

Ready

When the indication is showing 0% signal and the 

 green indication LED is

burning continuously again, then the zeroing action was successful.

Measurement check

For a controller: send a setpoint to the 

mini CORI-FLOW™ ML120

 and check the measured value.

Make sure the 

mini CORI-FLOW™ ML120

 indicates 0% at zero flow.

Your 

mini CORI-FLOW™ ML120

 is now ready for operation.

Bronkhorst High-Tech B.V., Nijverheidsstraat 1a, NL-7261AK Ruurlo  The Netherlands

T

 +31(0)573 45 88 00   

F

 +31(0)573 45 88 08   

I

 www.bronkhorst.com   

E

 info@bronkhorst.com

Summary of Contents for mini CORI-FLOW ML120

Page 1: ...Doc nr 9 17 093C Date 13 05 2016 Quick Installation Guide ...

Page 2: ...nterface Fieldbus interface description Manual Hookup diagram RS232 Interface with FLOW BUS protocol 9 17 027 9 16 132 FLOW BUS Interface 9 17 024 9 16 133 PROFIBUS DP Interface 9 17 025 9 16 134 DeviceNet Interface 9 17 026 9 16 135 Modbus interface 9 17 035 9 16 136 EtherCAT interface 9 17 063 9 16 137 PROFINET 9 17 095 9 16 146 i www These documents can be downloaded from the website http www b...

Page 3: ...ecommends performing an additional adequate drying procedure for those applications where remaining water particles may cause undesired reactions such as corrosion 4 1 Mount install instrument properly Install the mini CORI FLOW ML120 Meter Controller in the line and tighten the fittings according to the instructions of the supplier of the fittings During operation avoid external vibrations and sh...

Page 4: ...or leaks before applying fluid pressure Especially if toxic explosive or other dangerous fluids are used Avoid condensation due to cold liquids gases at high humidity environments Liquid tight is not the same as gas tight Please note that connections which are tight for liquid could still be untight for gas This can result in Air enclosure in the liquid e g when using external valves which might l...

Page 5: ...l the Air gas in the tubes for liquids or to remove all possible condensation drops for gases Purging during start up In order to remove condensation drops during start up flushing with dry gas for some minutes with high flow rate while tapping the valve is recommended Special control mode for purging In case of purging of a mini CORI FLOW ML120 controller give setpoint 100 to control the valve or...

Page 6: ... and the green LED will blink fast The procedure will take approx 45 seconds 45 sec When the signal is not stable the procedure will take longer max 6 retries will be performed The red LED will blink when signal is unstable noisy 10 5 Ready When the indication is showing 0 signal and the green indication LED is burning continuously again then the zeroing action was successful Measurement check For...

Reviews: