Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
42
Changing from digital operation to analog operation:
1. Set parameter
Init Reset
to 64 (unlocked)
2. Read parameter
IO Status
3. Add 64 to the read value
4. Write the new value to parameter
IO Status
5. Set parameter
Init Reset
to 82 (locked)
Changing from analog operation to digital operation:
1. Set parameter
Init Reset
to 64 (unlocked)
2. Read parameter
IO Status
3. Subtract 64 from the read value
4. Write the new value to parameter
IO Status
5. Set parameter
Init Reset
to 82 (locked)
Do not use this procedure if one of the following customized I/O options is installed:
·
C5S (digital output, enabled by setpoint)
·
I3C (digital input, controller mode valve close)
·
I8C (digital input, controller mode valve purge)
4.8
Fluid set
4.8.1
Basic fluid set parameters
For changing fluid set, flow range or operating conditions, using the FlowTune software is strongly advised. FlowTune
checks any changes for compatibility of the process conditions with the instrument. When the parameters discussed in this
section are changed manually, no such checks are performed, and the instrument output may become disordered or the
instrument may even be damaged if used in conditions the instrument is not suited for. Consult Bronkhorst service
personnel when in doubt.
Fluid Set Index
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…7
24
1/16
0x0030/49
With this parameter, any of the pre-configured fluid sets (maximum 8) can be selected. Each fluid set has its specific
(configurable) properties, such as
Fluid Name
,
Capacity
, etc.
Default value: 0 (fluid 1).
Fluid Name
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[10]
RW
-
25
1/17
0x8188…0x818C/33161…33165
This parameter contains the fluid name of the selected
Fluid Set Index
. This parameter may contain one of three value types:
·
Gas name, e.g. 'N2', 'He', 'C3H6 #2'.
·
CAS Registry Number, e.g. '7727-37-9', '7440-59-7', '115-07-1'
·
When
Fluid Name
= 'Mix', the instrument will use the gas mixture properties stored in parameters
Mix Fraction Type
,
Mix
Volume Temperature
,
Mix Volume Pressure
,
Mix Component Index
,
Mix Component Fraction
, and
Mix Component Fluid Name
.
Custom (fixed) fluid properties stored in the parameters
Density
,
Heat Capacity
,
Thermal Conductivity
, and
Viscosity
will be
ignored.
When no valid
Fluid Name
is entered or when insufficient or invalid fluid properties are entered, the instrument will be
forced into
until a valid fluid set configuration is made.
Capacity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
1E-10…
1E+10
21
1/13
0x8168…0x8169/33129…33130
This parameter sets the maximum readout/control value (100%) for the active
Fluid Set Index
in readout units corresponding
to
Capacity Unit
.
Capacity
is scaled when
Pressure Inlet
,
Fluid Temperature
or
Fluid Name
(or any of the mixture parameters) is
changed for the active fluid set.